Material testing equipment configured for your test method and specimen.

DERUI manufactures material testing equipment for quality control and R&D.
WhatsApp Consultation:+86 135 2795 8363

DR-C320 Differential Scanning Calorimeter

Model: DR-C320

Price: FOB, CIF

Payment:L/C, D/P, T/T

Package:Each set protected with resin fiber and PP film, then put into Strong wooden case with operation manual and video inside.

Delivery:15 days after order.

Safe and reliable
well-made
customization
Quality assurance
MessageWhatsApp
Hotline number +8615580327593
  • Product Details
  • Technical parameters
  • FAQ
  • Contact us
CONTROLLED HEAT-FLOW MEASUREMENT FOR POLYMERS

DR-C320 Differential Scanning Calorimeter

The DR-C320 records differential heat-flow response as a sample and reference follow a programmed temperature profile. Properly configured methods can identify glass-transition behaviour, melting and crystallization temperatures, transition enthalpies and selected curing or oxidation-related events.

Quick answer: choose a DSC from the required event and temperature program—not from maximum temperature alone. Define the sample, expected transition, heating/cooling range and rate, purge gas, pan type, calibration materials and calculation/reporting method before ordering.

MODELDR-C320
PUBLISHED RANGEAmbient to 500°C
HEATING RATE1–80°C/min
OPERATIONTouchscreen / PC
DERUI DR-C320 differential scanning calorimeter for polymer thermal analysis
Confirm the furnace and sensor configuration, cooling requirement, purge-gas connections, pan/crucible system, calibration kit, software outputs and usable measurement range for the planned DSC method.
READ THE CURVE BY EVENT TYPE

What a DSC Curve Can Show

Glass transition

A baseline step can be evaluated for Tg and step height using a defined construction and thermal history.

Melting and crystallization

Peaks can provide onset/peak temperatures and integrated enthalpy under controlled heating or cooling.

Curing or reaction

An exothermic or endothermic event can support reaction-temperature and enthalpy analysis when the method and pan are suitable.

Interpretation boundary: endothermic/exothermic direction depends on instrument and software convention. Crystallinity requires a justified reference enthalpy and corrections; it is not a direct number produced without material assumptions.
THE TEMPERATURE PROGRAM IS PART OF THE RESULT

Build a Reproducible DSC Method

A comparable curve requires control of sampling, pan, atmosphere, calibration, thermal history and data treatment—not only the same instrument model.

Define the question

Select Tg, melting, crystallization, cure, oxidation-related response or heat capacity and the governing method.

Prepare sample and pan

Control representative sampling, mass, contact, lid/seal, reference pan and volatile or reactive behaviour.

Calibrate and program

Verify temperature and heat-flow/enthalpy response, then set atmosphere, rate, isotherms and cycles.

Run and evaluate

Record raw curve, baseline, onset/peak/step construction, integration limits, sample mass and all calculations.

Thermal-history rule: first heating can reflect processing and storage history. A controlled cooling and second heating may answer a different material question; report the cycle used for every result.

SELECT THE PART THAT MATCHES THE REQUIRED RESULT

DSC Standards and Method Pathways

Method Reported result DR-C320 configuration review
ISO 11357-1:2023 General DSC principles, apparatus, sampling, calibration, procedure and report framework. Use as the base, then select the relevant subsequent part for the actual property.
ISO 11357-2:2020 Glass-transition temperature and associated step height for amorphous or partially crystalline plastics. Confirm baseline stability, temperature program, cooling and software constructions.
ISO 11357-3:2025 Melting/crystallization temperatures and enthalpies of crystalline or partially crystalline plastics. Confirm range, rate, atmosphere, calibration, integration and thermal-history cycle.
ASTM E1356 Glass-transition temperatures by DSC or DTA. Confirm current edition, sample, scan program and reported Tg construction.
ASTM E1269-24 Specific heat capacity by DSC for applicable thermally stable solids and liquids. Requires the complete comparative method, reference material and Cp workflow—not a generic DSC scan.

Compatibility rule: citing an ISO or ASTM number is not enough. The ordered temperature range, cooling, atmosphere, pans, calibration materials, software and verification evidence must support that method.

THERMAL ANALYSIS INSTRUMENTS ANSWER DIFFERENT QUESTIONS

Choose DSC for Heat-Flow Events

A thermal event in a DSC curve does not automatically identify decomposition, mechanical softening or processing flow.

DSC · DR-C320

Measures differential heat-flow response to a controlled temperature program.

TGA

Measures mass change versus temperature/time; use it to investigate decomposition or volatile loss directly.

DMA / rheology

Measures mechanical or viscoelastic response. Rubber cure behaviour belongs to a dedicated rheometer method.

Review rubber rheometer →

SMALL SIGNALS REQUIRE CONTROLLED PRACTICE

Control the Main Sources of Invalid DSC Data

Non-representative sample

Control location, particle size, homogeneity, mass and contact with the pan.

Wrong pan or seal

Volatile, reactive or pressure-generating samples require a compatible pan and approved handling.

Calibration mismatch

Temperature and enthalpy calibration must cover the actual range, rate and atmosphere where applicable.

Atmosphere error

Purge-gas identity, flow, dryness and switching can change baselines and reaction pathways.

Thermal lag

Sample mass and scan rate can shift peaks or distort separated events; keep method conditions consistent.

Baseline over-processing

Document blank subtraction, smoothing, baseline choice and integration limits; preserve raw data.

INPUTS FOR A COMPLETE DSC PROPOSAL

Build a Verifiable DR-C320 RFQ

Send a representative sample and expected thermal event so the range, pans, atmosphere, cooling and software can be reviewed together.

Material and decision

  • Polymer, blend, elastomer or thermoset
  • Expected transition/reaction
  • Required output and acceptance limit
  • Standard and current edition

Temperature program

  • Start/end temperatures
  • Heating and cooling rates
  • Isothermal segments and cycles
  • Required sub-ambient cooling

Sample environment

  • Sample mass and preparation
  • Pan/lid/sealing requirement
  • Purge or reactive gas
  • Volatile/pressure/safety concerns

Data and evidence

  • Calibration materials and intervals
  • Raw curves and calculations
  • FAT sample and acceptance checks
  • Power, computer, manuals and training
BUYER AND AI-READY ANSWERS

DR-C320 DSC FAQs

01What does a differential scanning calorimeter measure?
A DSC records differential heat-flow response between a sample and reference during a controlled temperature program, revealing method-defined transitions and reactions.
02Can the DR-C320 measure glass transition temperature?
It is intended for DSC thermal analysis, but Tg capability must be confirmed against the expected step size, baseline performance, temperature program, cooling and selected method.
03Does DSC directly measure polymer decomposition?
No. DSC detects heat-flow events. TGA directly measures mass change and is normally the clearer method for decomposition or volatile-loss behaviour.
04Why can first and second heating give different results?
First heating can include moulding, cooling, aging or storage history. Controlled cooling and second heating can reduce that history, but they answer a different question and must be reported separately.
05Can DSC calculate crystallinity automatically?
Software can integrate melting/crystallization enthalpy, but percent crystallinity requires an appropriate reference enthalpy and corrections for fillers, cold crystallization and sample composition.
06What is needed for a DR-C320 proposal?
Provide the material, method, expected event, temperature program, cooling, atmosphere, pan type, sample safety, outputs, calibration, FAT and reporting requirements.

Define the complete DSC method: sample + event + range + rate + atmosphere + pan + calibration + calculation.

Request a DR-C320 Proposal →

FacebookLinkedInXRedditWhatsApp

Temperature

Room temperature - 500℃

Temperature resolution

0.1 ℃

Heating rate

1 ~ 80 ℃ / min  

DSC range

0 ~ + 300 mw

DSC resolution

0.01 mW

DSC sensitivity

0.1 mW

Operation

Touch screen / PC

Power

Single phase, AC 220V, 50Hz

Accessory

One set computer& software

Q1: What is a DSC Differential Scanning Calorimeter?
A: A DSC measures heat flow associated with material phase transitions (melting, crystallization, glass transition) as a function of temperature or time, providing critical data for material characterization and quality control.

Q2: How does a DSC calorimeter work?
A: It compares heat flow between a sample and reference pan while heating, cooling, or holding temperature, detecting endothermic/exothermic events like melting, curing, or decomposition.

Q3: What is the ASTM standard for DSC testing?
A: Key standards include ASTM E967 (temperature calibration), ASTM E968 (heat flow calibration), and ASTM D3418 for polymer melting/crystallization measurements.

Q4: DSC instrument price range?
A: Prices range from $20,000 to $80,000+, depending on temperature range, sensitivity, automation, and software capabilities for data analysis.

Q5: How to calibrate a DSC instrument?
A: Use certified reference materials (e.g., indium, zinc) for temperature and enthalpy calibration. Verify baseline stability and purge gas flow regularly.

Q6: Why is DSC important in material science?
A: It characterizes polymer curing, purity, glass transitions, oxidation stability, and phase changes, essential for R&D, formulation, and quality assurance.

Q7: Applications of DSC analysis?
A: Used in pharmaceuticals, polymers, foods, chemicals, and metals to study melting, crystallization, glass transitions, specific heat, and thermal stability.

Q8: What is the difference between DSC and TGA?
A: DSC measures heat flow from phase changes. TGA (Thermogravimetric Analysis) measures mass loss due to decomposition, evaporation, or oxidation.

Q9: How to interpret a DSC curve?
A: Endothermic peaks (e.g., melting) absorb heat; exothermic peaks (e.g., crystallization) release heat. Glass transitions appear as baseline shifts.

Q10: What is glass transition temperature (Tg)?
A: Tg is the temperature where an amorphous material transitions from brittle to rubbery, detected as a step change in heat flow on a DSC curve.

Q11: What is the ISO standard for DSC?
A: ISO 11357 specifies methods for thermal analysis of plastics using DSC, including melting, crystallization, and oxidation induction time tests.

Q12: How to prepare samples for DSC?
A: Use small, homogeneous samples (1–10 mg) in sealed pans compatible with temperature range. Avoid overfilling to ensure thermal contact.

Q13: What safety features are essential?
A: Auto-cooling, pressure-resistant cells, emergency stop buttons, and temperature monitors to prevent overheating or sample decomposition risks.

Q14: Where to buy a DSC instrument?
A: Purchase from certified manufacturers or distributors offering installation, training, calibration services, and technical support for lab integration.

Q15: What is modulated DSC (MDSC)?
A: MDSC applies a modulated temperature waveform to separate reversible (heat capacity) and non-reversible (kinetic) thermal events for enhanced analysis.

Q16: How to choose a DSC instrument?
A: Consider temperature range (-150°C to 600°C+), sensitivity, cooling options, automation, and compliance with ASTM/ISO standards for your applications.

Q17: Can DSC test pharmaceuticals?
A: Yes, it analyzes drug polymorphism, purity, excipient compatibility, and amorphous content, critical for stability and formulation development.

Q18: What purge gases are used in DSC?
A: Nitrogen (inert), argon (high-temperature), and air or oxygen (oxidation studies) are common purge gases to control sample environment.

Q19: Why is DSC used in polymer testing?
A: It determines melting points, crystallization behavior, degree of cure, glass transitions, and thermal stability for polymer processing and application.

WhatsAppWhatsApp
WhatsApp
MessageMessage

Send Inquiry