What kind of foundation or lab space is needed for installation?

Date: January 2, 2026Views: 7857

Installation is more involved than for a benchtop unit. Key requirements include:

  • Substantial Foundation: A thick, reinforced concrete floor (often >12 inches) is mandatory to absorb the massive reaction forces and prevent vibration.

  • Adequate Space: Plan for the machine’s footprint plus 3-4 feet of clearance on all sides for operator access, fixture handling, and maintenance.

  • Utilities: Requires a 3-phase electrical supply for the hydraulic power unit and potentially process water cooling. The lab must have adequate ventilation and a stable ambient temperature.

  • Professional Installation: Always requires factory-trained technicians for assembly, alignment, and commissioning.